Sign in →

Test Code 028 Fibrinogen

Additional Codes

ALAB: FIB

PowerChart: Fibrinogen Activity

MA Cerner: Fibrinogen Level.

Specimen Requirement

BD sodium citrate blue-top tube filled to or above the etched line on the tube.

 

Greiner VACUETTE sodium citrate blue-top tube filled to the area within the indicator arrow printed on the side of the tube.

Additional Specimen Collection Information

Invert the sodium citrate tube end to end 4 times to thoroughly mix blood and anticoagulant. The tube should be left stoppered until time of testing. Specimens stored at room temperature should be centrifuged and tested within four hours of specimen collection. If the testing is not completed within 4 hours the plasma should be removed from the cells and transferred to a plastic tube. Freeze the sample for up to two weeks. Frozen samples should be rapidly thawed at 37°c while gently mixed. Thawed samples should be centrifuged and tested immediately. If testing cannot be performed immediately the sample may be held for a maximum of 2 hours refrigerated, until tested.

  • Precautions:

    • It is not recommended to use a winged blood collection set (Butterfly).
    • If a vacuum tube butterfly must be used, the first tube in the series will be under-filled. Collect a “discard citrate tube”, and then collect the citrate tube used for testing. This will ensure the proper blood to anticoagulant ratio.
    • If using a syringe, use a BD Blood Transfer Device to transfer blood from the syringe to the citrate tube within one minute of collection. Hold the syringe with the plunger in the up-right position and allow the vacuum present to fill the tube. Before collection, be sure the syringe plunger is pushed to the top of the syringe. The transfer device may pop up and needs to be held down to completely fill tube to between fill lines.
    • Do not “force fill” blood into a vacuum tube. This may result in activation of the platelets possibly leading to erroneous results. Allow the vacuum present to fill the tube.
    • Do not remove vacuum tube cap to add blood.
    • Specimens should be obtained from a single venipuncture with minimal tissue trauma. Never combine multiple blood collections into a single vacuum tube.
    • Never re-use a syringe that was used for a pic line saline flush.

Refer to the "Citrated Plasma for Coagulation Studies" protocol in the Quality Assurance Manual.

Performance

Testing is performed Monday through Sunday.

Routine orders: Typically completed within 4 hours; Mercy DV and MA West within 24 hours

Expedite orders: Typically completed within 1 hour after the specimen arrives at the testing site.

STAT orders: Typically completed within 30 minutes after the specimen arrives at the testing site.

Reference Values

200-393 mg/dL

  • Critical value: < 80 mg/dL

CPT Code Information

85384 Coagulopathy Testing: Fibrinogen Activity

Useful For

Detecting increased or decreased fibrinogen (factor I) concentration of acquired or congenital origin

Monitoring severity and treatment of disseminated intravascular coagulation and fibrinolysis

LOINC Code Information

3255-7 Fibrinogen [Mass/Volume] In Platelet Poor Plasma By Coagulation Assay

Test Method

Clot Based Assay; ACL Systems